Perform every time CPU has been replaced, or calibration chip removed.
From Main Menu hold "Alt+C" to enter calibration screen. To enter password hold: "Ctrl+D"
"Ctrl+A" "Ctrl+D"Press L to Load ( Do not save to CPU - choose "NO" Press the number of your choice "Yes" Hold "Esc" to exit calibration screen From Main Menu upload Gen1.010 and download to Press.
To copy calibration files on the floppy type from DOS screen: disk copy c*.*clb a:
CALSEM - Oil temperature alarm disable/enable
To get to calibration screen hold "Alt + C" Enter password - Depress "Ctrl" while typing "DAD" and "Enter" or hold "Ctrl+D", "Ctrl+A", "Ctrl+D" From the list pick (item 13) - max oil temp. (note value -600) To disable oil temp. alarm change value to -3200 Hit "Enter"
|